Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
12. Place the vial in the sample chamber making sure that
the marks are aligned.
13. Press
TEST
key.
Wait for a
reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
See page 37.
*,** mg/l free Cl
*,** mg/l comb. Cl
*,** mg/l total Cl
T1 accepted
prepare T2
press TEST
Countdown
3:00