2
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
Press the appropriate numerical key to select the required
resolution, e.g.: [Shift] + [3] for 0.01.
Note:
Please enter the required resolution according to the
instrument pre-sets:
range
max. resolutions
0.000 ...9.999
0.001
10.00 ...99.99
0.01
100.0... 999.9
0.1
1000 ...9999
1
Measurement procedure with standards of
known concentration:
The display shows:
Prepare Zero and press [Zero] key.
Note:
Use deionized water or reagent blank value.
The display shows:
Enter the concentration of the first standard;
e.g.: [Shift] + [0][.][0][5]
• One step back with [ESC].
• Press [F1] key to reset numerical input.
Confirm with [ ] key.
The display shows:
Prepare the first standard and press [Test] key.
The display shows the input value and the measured
absorption value. Confirm with [ ] key.
Enter the concentration of the second standard;
e.g.: [Shift] + [0][.][1]
• One step back with [ESC].
• Press [F1] key to reset numerical input.
Confirm with [ ] key.
< User concentr.>
prepare Zero
press ZERO
choose resolution
1: 1
2: 0.1
3: 0.01
4: 0.001
< User concentr.>
S1: 0.05 mg/l
prepare
press TEST
< User concentr.>
Zero accepted
S1: +________
| ESC | F1
S1 accepted
S2: +________
| ESC | F1
S1: 0.05 mg/l
mAbs: 12
3
0
0
5
.
0
1
.
Test
Zero
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......