135
Medfusion® Model 3500 pump V6
Operator’s Manual
Prompt to Prime
Defines if a pump user will be prompted to
Prime
as a last step in setting up an
infusion. Select from two values—
Enabled
or
Disabled
.
Re-enter Weight
Defines if a pump user will be prompted to re-enter a patient’s weight to
confirm the setting (a double check). Select from two values—
Enabled
or
Disabled
.
Standby Time
Defines if a pump user will be prompted to enter a
Standby
time during
infusion setup. Select from two values—
Enabled
or
Disabled
.
Volume Limit
Defines if a pump user will be prompted to enter a
Volume Limit
during
infusion setup. Select from two values—
Enabled
or
Disabled
.
Program Options
Program Options are optional settings that affect the
current
infusion initiated in the defined profile. While
individual options can be enabled for a profile, they can
also be disabled in individual drug programs within
that profile. Settings in drug programs take priority over
those of the profile.
Bolus Dose
Defines if a pump user will be able to enter a
Bolus Dose
during infusion setup.
Delayed Start
Defines if a pump user will be able to enter a
Delayed
Start time during
infusion setup. The range is 0 to 6 hours.
KVO
Defines if a pump user will be able to enter a Keep-Vein-Open (KVO) rate.
This value cannot be set or changed while an infusion is running.
Loading Dose
Defines if a pump user will be able to enter a
Loading Dose
during infusion
setup.
Override Alarm Loudness
Defines if a pump user will be able to change the loudness level of a pump’s
alarm during an infusion.
Override Occlusion Limit
Defines if a pump user will be able to override the profile default setting of
the pressure at which the pump will consider the line occluded for the current
infusion.
Periodic Callback Alarm
Defines if a pump user will be able to enter a
Periodic Callback Alarm
time
during infusion setup. The range is 0 to 8 hours.
Standby
Defines if a pump user will be able to enter a
Standby
time during infusion
setup. The range is 0 to 24 hours.
Toggle Vol Empty Tone
Defines if a pump user will be able to enable the
Syringe Vol Empty Alarm
Tone
for the
current
infusion.
Toggle FlowSentry™
Defines if a pump user will be able to disable the
FlowSentry™
pressure monitor
(for rapid occlusion detection) feature for the current infusion.
Toggle Near Empty Tone
Defines if a pump user will be able to disable the
Syringe Near Empty Alarm
Tone
for the
current
infusion.
Toggle PVD/PDD
Displayed on the pump options menu as “Change to Dose” or “Change to
Volume.” Defines if a pump user will be able to change the profile default setting
to display the delivered
Program Dose
or the delivered
Program Volume
.