85
Medfusion® Model 3500 pump V6
Operator’s Manual
Disable/enable FlowSentry™
pressure monitoring
The pump has an internal program for rapid
occlusion detection called FlowSentry™ pressure
monitoring. The pump may have been configured
to either turn this feature on or off. To change
the default setting for a specific infusion, use this
option.
Note:
With certain syringes, such as those for
enteral delivery, FlowSentry™ pressure monitoring
is automatically disabled and cannot be enabled.
1. With the pump running or stopped, press
Options
.
2. Use the
number
keys to choose
Disable
FlowSentry
or
Enable FlowSentry
(only one of the above will be displayed,
depending on whether the rapid occlusion
detection feature is currently active or not).
(You may have to press
More
to see this
option.)
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
BEGINNING
1
DISABLE F
LOW
S
ENTRY
2
DISABLE NEAR EMPTY TONE
3
DISABLE VOL EMPTY TONE
Disable/enable near empty alarm
tone
The pump may be configured to have an alarm
sound to indicate when a syringe is
nearly
empty.
(Alarm triggering is calculated based on the syringe
size used and the volume initially calculated; see
page 102 for an explanation of this alarm.) Use
this option to enable or disable this feature for this
infusion.
1. With the pump running or stopped, press
Options
.
2. Use the
number
keys to choose
Disable
Near Empty Tone
or
Enable Near
Empty Tone
(only one of the above will be
displayed, depending on whether the Near
Empty Alarm feature is currently active or
not). (You may have to press
More
to see
this option.)
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
BEGINNING
1
DISABLE F
LOW
S
ENTRY
2
DISABLE NEAR EMPTY TONE
3
DISABLE VOL EMPTY TONE
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
Summary of Contents for Medfusion 3500