CONFIRM DRUG PROGRAM SELECTION?
HEM/ONC
NO
YES
PRESS THE NUMBER TO SELECT
MORE
1
m
L/
Hr
5
RECALL LAST SETTINGS
2
VOLUME/TIME
3
DOSE/TIME
4
DOSE / KG / TIME
The pump requires a confirmation for drug
program selections; press
Yes
to confirm the
selection, or press
No
to return to the
Select
screen.
If “manual” modes are available in the custom
Configuration and selected for an infusion, the
pump prompts for a confirmation as well. Press
Yes
to confirm the selection, or press
No
to return
to the
Select
screen.
After a drug program selection is made the pump
will prompt to select the syringe model and size
before programming the selected mode (see page
33).
Specific information for programming the various
delivery types follows the instructions for the
syringe.
Select a Delivery Mode (Medfusion®
Standard Configuration)
1 If the pump does not have a custom
Configuration, the Medfusion
®
Standard
Configuration menu displays.
2. Select a delivery mode by pressing its
number
. If the desired mode is not
displayed, press
More
to display additional
delivery modes (if present).
3. The pump requires confirmation of manual
mode selection. Press
Yes
to confirm the
selection, or press
No
to return to the
Select
screen.
After a delivery mode selection is made the pump
will prompt to select the syringe model and size
before programming the selected mode (see page
33).
Specific information for programming the various
delivery types follows the instructions for the
syringe.
m
L /
Hr
CONFIRM MANUAL PROGRAMMING / DELIVERY MODE?
NO
YES
m
L /
Hr
CONFIRM MANUAL PROGRAMMING / DELIVERY MODE?
NO
YES