107
Medfusion® Model 3500 pump V6
Operator’s Manual
Alarm Message
Priority
Remedy
Restricted Flow - Bolus
Cancelled
Medium
Bolus delivery was cancelled before completion. While the pump
was delivering a bolus, the flow rate was restricted and initiated rate
reduction, which eventually fell below the main infusion rate or
extended past the maximum bolus time. Check the infusion line for
occlusion.
Restricted Flow - Loading
Cancelled
Medium
Loading dose delivery was cancelled before completion. While the pump
was delivering a loading dose, the flow rate was restricted and initiated
rate reduction, which eventually fell below the main infusion rate or
extended past the maximum loading dose time. Check the infusion line
for occlusion.
Restricted Flow - Rate
Reduced
Medium
If the bolus or loading dose rate is high enough to trigger an occlusion
alarm, the software reduces the rate until the bolus/loading dose is
successfully delivered or, if the rate falls below a certain level or takes
longer than the pump allows, is cancelled. With each rate reduction,
this process repeats. Once silenced, the audible alarm tone will not
reactivate, even if the process repeats.
Set Volume Limit Before
KVO
Limit
Programming of a KVO rate was attempted, but there is no volume limit
set. Program a volume limit, then set a KVO rate.
Syringe Does Not Match
Entry
Limit
The drug program indicates a specific syringe, and the syringe in use is
different. Verify the correct drug program and syringe are being used.
Syringe Empty
High
Press
í
. The syringe is now empty. Either:
• Prepare and load another syringe.
• Return to the
Select the Profile
screen, where no alarms
sound.
• Turn the pump off.
Syringe Empty - Stop
High
Press
í
. The pump has either calculated (based on the syringe model
in use) that the syringe should be at its forward-most travel point, or the
pump’s Syringe Plunger Driver has moved as far forward as is possible.
Either:
• Prepare and load another syringe.
• Return to the
Select the Profile
screen, where no alarms
sound.
• Turn the pump off.
Syringe Flange Not in
Place
High
Press
í
. The syringe is not loaded correctly, or the sensor has failed.
Make sure the syringe flange is properly inserted into the flange clip.
Proceed with the infusion, if possible. If the alarm recurs, remove the
pump from service for repair by a trained biomedical service technician.
Syringe Near Empty
Medium
or Low*
The pump is indicating the syringe will soon be empty. The “Near
Empty Time” interval ranges from 1 - 240 minutes. This time interval
may be set via the “Near Empty Time” in PharmGuard
®
Toolbox 2, as
well as the alarm priority,
Low
or
Medium
(the default priority is
Low
).
The alarm tone may be enabled/disabled in the
Options
menu if the
Option is made available in the Profile; the default is enabled.
Summary of Contents for Medfusion 3500
Page 151: ......