44
Medfusion® Model 3500 pump V6
Operator’s Manual
OPTIONS
CLEAR TOTALS
PRESS < « > TO BEGIN INFUSION
MAIN MENU
CONC
0.6
mg
/
m
L
PVD
0
m
L
DOSE
5
MG
TIME
02:15:00
TIME REMAINING
02:15:00
RATE
3.7
m
L/
Hr
Main Menu
– returns to the
M
ain
M
enu
.
Options
– allows selection from a list of variables (e.g.,
Standby, etc.).
Clear Totals
– clears total volume delivered, and PVD or
PDD
.
7. Confirm the settings and start the infusion,
or use the three menu softkeys shown at
right to modify the settings.
8. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
9. When all delivery settings are confirmed as
correct, press
É
to begin infusion.