Alarm Message
Priority
Remedy
Check Syringe Plunger
Sensor
High
Sensor is active when syringe is not loaded correctly. Check the syringe
plunger holders. Ensure the syringe barrel clamp, flange holder, and
plunger holders are properly engaged and the holders move freely.
If they do not operate correctly or if the alarm persists, immediately
remove the pump from service for repair by a trained biomedical service
technician.
Data Does Not Match
Previous Entry
Limit
In cases where confirmation of entered data is required, the
confirmation value did not match the originally programmed value. Re-
enter the correct value.
Enter time as HH:MM
(MM:SS) - Use 01:00 for 60
minutes (seconds)
Limit
The time value entered is greater than the maximum allowed for the
minutes or seconds. Please try a lower value, and use 01:00 for a value
greater than 59 minutes when programming hours, and 59 seconds
when programming minutes.
Force Sensor Bgnd Test
High
A background self-test was performed on the force sensor while running
an infusion, and the test failed due to a sensor malfunction. Press
í
.
Press
É
to restart the infusion. If the alarm recurs, remove the pump
from service for repair by a trained biomedical service technician.
Infusion Complete
High
The programmed infusion (based upon the values entered in the pump)
has been delivered. Press
í
. Either begin a new infusion, return to the
Select The Profile
screen (where no alarms sound), or turn off the
pump.
Input Out of Range -
Greater Than Max Value
Limit
The value entered is greater than the maximum allowed for the item.
Please try a lower value.
Input Out of Range - Less
Than Min Value
Limit
The value entered is less than the minimum allowed for the item. Please
try a higher value.
Invalid Entry
Limit
There is a problem with the drug program selected, and the pump
cannot deliver it. Contact your PharmGuard
®
Toolbox 2 Administrator,
who will have to review and correct the problem.
Invalid Infusion Parameter
Combination
Low
The combination of infusion parameters does not allow a valid flow rate
to be calculated. Verify all infusion settings. If the invalid combination
is part of a Drug Program, contact your PharmGuard
®
Toolbox 2
Administrator, who will have to review and correct the problem.
Invalid Syringe Size
High
Syringe size not recognized for the model selected, or the syringe barrel
clamp was lifted during delivery. Review syringe sizes in the technical
specifications section, and do not lift the syringe barrel clamp during
delivery. If the alarm persists, remove the pump from service for repair
by a trained biomedical technician.
KVO in Progress
Medium
Informational message that volume limit has been reached and KVO
infusion is in progress. Press
í
.
Limits Adjusted - Cannot
Deliver All Doses
Limit
The PharmGuard® Toolbox 2 specified soft limit values and syringe
specified during programming would allow values to be entered that
result in a rate outside the allowable limits. The pump has adjusted the
limits to keep the rate inside allowable limits. Verify the values entered
and the syringe size.