Making changes during delivery
At right is an illustration of a delivery screen. There
are eight options available during any standard
infusion. While running an infusion, you may:
• Lock/Unlock the keypad.
• Change the Dosage Rate.
• Select certain Program Options.
• Clear TVD
• Press
è
to set, change, or deliver a bolus.
• Press
á
to pause or end infusion delivery.
• Press
ë
to begin pump shutdown.
• Press
í
to silence an alarm.
Changing delivery rate
The Medfusion
®
Model 3500 pump V6 offers ways
of changing delivery rate either before beginning an
infusion, during delivery or when delivery is paused.
LOCK
N
CONC
0.5
mg
/
m
L
TVD
0.005
m
L
WEIGHT
5
KG
DOSE
6
MCG/KG/MIN
RATE
3.6
m
L/
Hr
CHG DOSE
OPTIONS
CLEAR TVD
DOBUT
amine
0.5
mg
/
m
L
PICU
Changing rate during delivery
The infusion rate can be changed during any
delivery session, except when using Intermittent
delivery modes.
1. From any delivery screen, press
CHG Rate
(or
CHG Dose
).
2. The
Enter Rate
screen appears.
Infusion
does not stop
. The current setting continues
to be delivered until the new rate is
entered
.
Note the current setting is always shown in
the top of the display.
3. Use the
number
keys to set the dose rate.
Rate is limited by the maximum and
minimum delivery rate for syringe size.
If a rate change is not desired, press
â
.
The delivery rate does not change until
Enter
is pressed.
4. When the desired rate is displayed press
Enter
to accept the new setting. (The new
rate is not effective until
Enter
is pressed.)
BACKSPACE
ENTER
ENTER RATE - PRESS ENTER TO CONTINUE
HIGH: 944
LOW: 0.1
CLEAR
m
L/
Hr
40
60
CURRENT SETTING