40
Medfusion® Model 3500 pump V6
Operator’s Manual
OPTIONS
CLEAR TOTALS
PRESS < « > TO BEGIN INFUSION
CHG DOSE
MAIN MENU
CONC
0.6
mg
/
m
L
PVD
0
m
L
DOSE
1.5
MCG/MIN
RATE
0.15
m
L/
Hr
Main Menu
– returns to the
M
ain
M
enu
.
Chg Dose
– allows modifying the set dosage without
starting from scratch.
Options
– allows selection from a list of variables (e.g.,
bolus, volume limit, etc.).
Clear Totals
– clears total volume delivered, and PVD or
PDD
.
6. Confirm the settings and start the infusion,
or use the four menu softkeys shown at right
to modify the settings.
7. Prior to connecting to the patient, press
è
to initiate the priming sequence (page 58).
8. If desired, program a Bolus Dose, Loading
Dose or Volume Limit (see Options, page
60).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
9. When all delivery settings are confirmed as
correct, press
É
to begin infusion.