Medfusion® Model 3500 pump V6
Operator’s Manual
The pump may be customized with an
additional patient weight/body surface
area confirmation screen. If so, re-enter the
patient weight/body surface area, then press
Enter
to accept the setting.
5. At the
Dose Rate
screen, press the
number
keys to set the dose. Range
limits are determined by installed syringe
size and/or the custom setup in the
Configuration. When the setting is correct
press
Enter
to accept the dose.
6. If desired (and if they are enabled),
program a Bolus Dose, Loading Dose or
Volume Limit (see Options, page 60).
7. At the
Begin Infusion
screen all
values are set and the pump pauses
(the pump may be configured with
additional confirmation screens, requiring
confirmation of all settings again before
beginning the infusion).
8. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
9. Confirm the settings and start the infusion,
or use the four menu softkeys shown at
right to modify the settings.
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
10. When all delivery settings are confirmed as
correct, press
É
to begin infusion.
BACKSPACE
ENTER
RE-ENTER PATIENT WEIGHT - PRESS ENTER TO CONTINUE
HIGH: 250
LOW: 0.25
CLEAR
KG
BACKSPACE
ENTER
ENTER DOSE RATE
-
PRESS ENTER TO CONTINUE
HIGH: 0.2012
LOW: 0.0001
CLEAR
MCG / KG / HR
0.2
OPTIONS
CLEAR TOTALS
PRESS < « > TO BEGIN INFUSION
CHG DOSE
MAIN MENU
CONC
0.25
mg
/
m
L
PVD
0
m
L
WEIGHT
10
kg
DOSE
0.2
mcg
/
kg
/
Hr
RATE
8
m
L/
Hr
Main Menu
– returns to the
M
ain
M
enu
.
Chg Dose
– allows modifying the set dosage without
starting from scratch.
Options
– allows selection from a list of variables (e.g.,
bolus, volume limit, etc.).
Clear Totals
– clears total volume delivered, and PVD or
PDD.