72
Medfusion® Model 3500 pump V6
Operator’s Manual
2. If, however, the loading dose rate reduces
below the infusion rate, or the dose time
increases past the maximum loading dose
time, the loading dose is
cancelled
. Delivery
stops.
• Press
í
to silence the alarm.
Determine the reason the loading dose
was unable to be delivered before re-
attempting a loading dose. Once the
reason for the cancellation is cleared,
loading dose delivery may be re-initiated
(if desired). Unless the loading dose
settings are reprogrammed, the last
loading dose rate setting from just
prior to cancellation will continue to be
applied. Reprogram the loading dose, if
desired.
• From the
Continue Same Load?
screen, to continue delivering the
loading dose at the new, reduced rate,
press
Yes
. To alter the rate, press
No
,
press
á
to stop the normal infusion,
then enter the
Options
menu to
reprogram the Loading Dose (this
method restarts the infusion from the
beginning, so you may need to manually
subtract any loading dose delivered prior
to the cancellation from the amount
programmed).
• If
Recall Last Settings
is used,
when
É
is pressed the
Begin
Loading Dose Delivery?
screen
appears with the new, reduced rate. If
desired, use the
CHG Dose
key to alter
the rate (for example, decrease loading
Dose and/or increase loading Time).
CONTINUE SAME LOAD?
BOLUS
5
mg
/
kg
DELIVERED
4.9690
m
L
REMAINING
08:28
MM:SS
99.3799
MG
TOTAL BOLUS
200
MG
RATE
70.2
m
L/
Hr
YES
NO
RESTRICTED FLOW - LOADING CANCELLED
a
minocaProic
a
cid
d
riP
20
mg
/
m
L
LOAD
5
mg
/
kg
DELIVERED
4.9690
m
L
REMAINING
08:28
MM:SS
99.3799
MG
TOTAL LOAD
200
MG
RATE
70.2
m
L/
Hr
CANCEL
CHG DOSE
NO
BEGIN LOADING DOSE DELIVERY?
LOAD
5
MG/KG
TIME
08:33
MM:SS
RATE
70.2
m
L/
Hr
YES
totaL
Loading
doSe
200
mg