Bolus Volume
The recommended minimum bolus volume is
1 mL or 10 percent of the syringe volume of the
syringe used, which ever is less. In other words, for
a 1 mL syringe (regardless of the actual volume of
medication contained), the minimum recommended
bolus volume is 0.1 mL. For a 60 mL syringe, the
recommended minimum bolus volume (regardless
of the actual volume of medication contained), is
1 mL. Bolus volumes delivered that are less than
the recommended minimum bolus volumes for the
syringe used may result in delivery inaccuracies.
Syringe Size
Minimum Recommended Bolus
Volume
1 mL
0.1 mL
3 mL
0.3 mL
5 mL
0.5 mL
6 mL
0.6 mL
10 mL
1 mL
12 mL
1 mL
20 mL
1 mL
30/35 mL
1 mL
50 mL
1 mL
60 mL
1 mL
Use appropriate internal
diameter tubing
For the most consistently precise control of all
fluids and medications, it is recommended to use
appropriate
internal diameter tubing
, as well as to
minimize residual volumes
between syringe and
patient by reducing:
a) Tubing internal diameter (generally, for rates
under 5 mL/hr small internal diameter tubing
is recommended; for rates 5 mL/hr or above,
tubing with a larger internal diameter should be
used to reduce occlusion alarms)
b) Tubing length
c) Size of in-line filters
d) Number of stopcocks
This makes the infusion setup less variable or
compliant, which in turn reduces the start time for
fluid to reach the patient – and reduces time required
for detecting an occlusion and maintains delivery
accuracy.
Use only the standard syringes specified on the
table on page 128 and supplied with PharmGuard
®
Toolbox 2, or supplied with the PharmGuard
®
Supported Syringes (PSS) Series 2 (available from
Smiths Medical). Smiths Medical also provides a
variety of infusion sets that can be used with the
pump. See www.smiths-medical.com for information
regarding infusion sets available from Smiths Medical,
or contact your sales representative.