122
Medfusion® Model 3500 pump V6
Operator’s Manual
Technical specifications
Specifications
Below are the technical specifications of the Medfusion
®
Model 3500 pump V6. Testing was performed using
the syringes listed in the table on page 128 and the following Medfusion
®
Administration Sets: 1) Model
536040 for low flow rates and 2) Model MX448HL60 for high flow rates.
Minimum increment of
resolution
0.01 mL/hr
Size (pump only)
27cm wide × 13cm high × 14cm deep (10.5” × 5.1” × 5.5”)
Weight
Approximately 1.7 kg (3.8 lbs.)
Classification
Type CF (protection from electric shock)
Moisture protection
IPX3 (sprayproof)
Pump Alarms
Syringe empty, syringe near empty, syringe not loaded properly, incorrect syringe
size, infusion complete, KVO, battery low/damaged/not working, incorrect
calculation, (value) out of range, occlusion/fluid restricted/rate reduced/
pressure increasing, pump stopped, system fault, system advisory (set time/date,
maintenance recommended, data corrupted)
For a complete list of alarms, see the General system alarms and alerts table
starting on page 104.
Time to occlusion alarm
Very High (50 psi) testing performed using B-D® 1 mL syringe; High and Very
Low testing performed using B-D® 20 mL syringe (syringes meet the requirements
of the ISO 7886-2 standard).
Flow Rate (mL/hr) Syringe (Mfg / Size) Occlusion Setting
Max time to
occlusion (Start-
up) hr:min:sec
1 (minimum)
B-D® 1 mL
Very High (50 psi)
00:10:00
B-D® 20 mL
High (16 psi)
00:31:00
Very Low (4 psi)
00:09:00
5 (intermediate)
B-D® 1 mL
Very High (50 psi)
00:01:00
B-D® 20 mL
High (16 psi)
00:06:00
Very Low (4 psi)
00:05:00
Power source
Internal battery:
Rechargeable (nickel metal hydride)
AC power:
100-240 VAC, 50/60 Hz, 30VA; Safety Class II
DC Power Supply
(optional)
External DC Input 6 – 14 VDC, 7 Watts Safety Class II – Type
CF (must meet IEC 60601-1 Medical Equipment Standards)
Note:
The DC Connector Port is functional only on pumps where the External DC
Jack Connection symbol (
Ö
) appears on the pump serial number label (located
on the bottom of the pump).
Battery Life
500 full discharge/charge cycles or 2 years, whichever comes first
Battery longevity
At 25° C (77°F), a fully charged battery will run the pump for at least 6 hours when
running a continuous rate of 5 mL/hr and using a 60mL syringe.