41
Medfusion® Model 3500 pump V6
Operator’s Manual
PRESS THE NUMBER TO SELECT THE CONCENTRATION UNITS
1
GRAMS
5
MILLIEQUIVALENTS
2
MG
6
MILLIMOLES
3
MCG
4
UNITS
BACKSPACE
ENTER
ENTER CONCENTRATION - PRESS ENTER TO CONTINUE
HIGH: 9999
LOW: 0.0001
CLEAR
mg
/
m
L
0.
9
PRESS THE NUMBER TO SELECT THE DELIVERY UNITS
1
MG
2
MCG
BACKSPACE
ENTER
ENTER DOSE RATE - PRESS ENTER TO CONTINUE
HIGH: 8732
LOW: 2.16
CLEAR
MG / DAY
5
Main Menu
– returns to the
M
ain
M
enu
.
Chg Dose
– allows modifying the set dosage without
starting from scratch.
Options
– allows selection from a list of variables (e.g.,
bolus, volume limit, etc.).
Clear Totals
– clears total volume delivered, and PVD or
PDD
.
Dose / Day
The Dose / Day mode may be available on a pump.
Program Dose / Day by setting:
•
Drug concentration units
(i.e., g/mL, mg/mL,
mcg/mL, mMol/mL, mEq/mL, or Units/mL)
•
Drug concentration
•
Drug delivery units
(i.e., g, mg, mcg, ng, mMol,
mEq, or Units)
•
Dose rate
•
Volume limit, KVO, loading dose, or bolus
dose may also be programmed (these options
are only available if they have been enabled in
the profile or specified in the Drug Program
using PharmGuard
®
Toolbox 2)
At the selection menu, choose
Dose/Day
(or choose
a drug program that uses the infusion type). Choose
the syringe model and size, and load the syringe as
instructed on page 33.
1. Use the
number
keys to select the
medication concentration
units
. (The actual
units available on a pump depends on the
custom setup.) Press
Enter
to accept the
medication concentration units.
2. Use the
number
keys to set medication
concentration. Concentration is limited to
4 digits. Press
Clear
to clear an entry. Press
Backspace
to erase a character. Press
Enter
to accept the setting.
3. Use the
number
keys to set the delivery
units. (The actual units available on a
pump depends on the custom setup.)
If
only one delivery unit is available for the
concentration units selected, this screen
will not appear
. Press
Enter
to accept the
delivery units.
4. Use the
number
keys to set dose rate. Range
limits for dose rate are determined by
syringe size and/or the custom setup in the
Configuration. The high and low rate limits
display on screen. When the dose is correct,
press
Enter
.
5. At the
Begin Infusion
screen all values
are set and the pump pauses (the pump may
be configured with additional confirmation
screens, requiring confirmation of all
settings again before beginning the
infusion).
6. Confirm the settings and start the infusion,
or use the four menu softkeys shown at
right to modify the settings.
OPTIONS
CLEAR TOTALS
PRESS < « > TO BEGIN INFUSION
CHG DOSE
MAIN MENU
CONC
0.9
mcg
/
m
L
PVD
0
m
L
DOSE
5
MG/DAY
RATE
0.232
m
L/
Hr
Summary of Contents for Medfusion 3500