Medfusion® Model 3500 pump V6
Operator’s Manual
Key
When pump is paused
When pump is delivering
Alarm Silence (
í
)
Silences audible alarm. Allows operator
to switch the display backlight from
bright to dim or dim to bright. Allows
redisplay of the text for the last active
alarm.
Silences audible alarm. Allows operator to
switch the display backlight from bright to
dim or dim to bright.
Power (
ë
)
Silence certain active alarms. Push and
hold to turn pump off.
Silence certain active alarms (some alarms
must be silenced with
í
before
ë
will
function). Push and release, then push and
hold to turn pump off.
Menu keys (softkeys)
Function is defined on the display.
Function is defined on the display.
Back (
â
)
Reverts to a previous step or level.
Reverts to a previous step or level if adjusting
settings. Briefly displays the battery life
indicator and PDD/PVD.
Numbers & Decimal
Set number values or selects menu
items.
Set number values or selects menu items.
Stop (
á
)
N/A
Stops delivery (pump remains on).
Start (
É
)
Starts delivery.
N/A
Prime / Bolus (
è
)
Begins priming after confirmation.
Displays the programmed bolus programming
or confirmation/
Begin Delivery
screen.
Indicator
What it means
ò
- Alarm
The
Alarm
indicators (yellow or red) are on whenever the pump is in an alarm
condition. The specific details of each alarm are covered in the Alarm section later in
this manual.
ô
- AC Line
The
AC Line
indicator (green) is on whenever the pump is connected to “mains” line
power. It is off when the pump is not connected to an active AC line.
õ
- Battery
The
Battery
indicator (green) blinks on & off whenever the pump is operating on
internal battery power, and remains on when battery is charging.
ö
- Lock
The
Lock
indicator indicates the pump has been locked into its current operational
mode.
While this indicator is lit, the keypad is locked and no changes can be made to settings.
Attempting to stop or change an infusion while locked will result in an alarm and an
informational message.
ù
- Infusing
The
Infusing
indicators are 3 green lights, which illuminate in sequence right to left
when the pump is infusing.
During intermittent volume over time delivery mode, a single Infusing indicator lights
during the time between infusions.
When the pump is off or stopped, the Infusing indicator is not lit.