Medfusion® Model 3500 pump V6
Operator’s Manual
Override alarm loudness
Occasionally it may be necessary to change the
loudness of the audible alarm. For instance, if the
infusion is taking place in an area with a lot of
ambient room noise the loudest alarm setting may
be wanted. The alarm loudness can be set from
Level 1 (quietest) to Level 5 (loudest), however it
cannot be disabled or completely silenced.
1. With the pump running or stopped, press
Options
.
2. Use the
number
keys to choose
Override
Alarm Loudness
. (You may have to
press
More
to see this option.)
3. The
highlight
shows the pump’s current
default setting. Use the
number
keys to
choose the desired alarm loudness setting,
then press
Enter
.
To determine which setting meets the
needs for the infusion, use the
number
key
to choose a loudness level then press
Test
.
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
STANDBY
5
DELAYED START
2
VOLUME LIMIT
6
LOADING DOSE
3
OVERRIDE OCCL LIMIT
7
BOLUS DOSE
4
OVERRIDE ALARM LOUDNESS
8
KVO
ADJUST LOUDNESS - PRESS ENTER TO CONTINUE
ENTER
TEST
1
LEVEL 1
5
LEVEL 5
2
LEVEL 2
3
LEVEL 3
4
LEVEL 4
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr