129
Medfusion® Model 3500 pump V6
Operator’s Manual
Flow rate & minimum volume by manufacturer & size - Standard syringes
For flow rate and minimum volume information
regarding syringes available in the PharmGuard
®
Supported Syringes (PSS) Series 2, see the product
literature provided with that product. Please see the
tables below for information regarding the lowest
recommended
rates for each syringe size used on the
pump.
The table below is provided for convenience. Should
changes to the listed syringes occur, the values shown
here may be superseded (see the
Instructions for Use
provided with the Medfusion
®
Standard Syringes,
Series 2.)
Model
(Manufacturer)
Size
Min. Rate
Max. Rate Minimum volume
to infuse
Minimum recommended rate
mL
mL/hr
mL/hr
mL
mL/hr
FlowSentry™
disabled
FlowSentry™
enabled
B-D® (Becton-Dickinson)
Tuberculin (Slip Tip)
1
0.01
29
0.0016
0.033
N/A*
Luer Lok™
1
0.01
29
0.0016
0.033
N/A*
3
0.01
98
0.005
0.1
0.03
5
0.03
191
0.0083
0.17
0.05
10
0.05
277
0.0166
0.33
0.1
20
0.1
483
0.0333
0.67
0.2
30
0.1
622
0.05
1
0.3
60
0.1
944
0.1
2
0.6
Monoject® (Covidien)
Tuberculin (Slip Tip)
1
0.01
29
0.0016
0.033
N/A*
Luer Lock
3
0.01
105
0.005
0.1
0.03
6
0.03
213
0.01
0.2
0.06
12
0.05
325
0.02
0.4
0.12
20
0.1
536
0.0333
0.67
0.2
35
0.1
735
0.0583
1.2
0.35
60
0.1
944
0.1
2
0.6
Terumo® (Terumo® Medical)
Tuberculin (Slip Tip)
1
0.01
29
0.0016
0.033
N/A*
Luer Lock
3
0.01
106
0.005
0.1
N/A*
5
0.03
225
0.0083
0.17
N/A*
10
0.05
333
0.0166
0.33
N/A*
20
0.1
541
0.0333
0.67
N/A*
30
0.1
712
0.05
1
N/A*
60
0.11
1130
0.1
2
N/A*
Summary of Contents for Medfusion 3500