Medfusion® Model 3500 pump V6
Operator’s Manual
On screen display of internal
battery power level gauge
Whenever the pump is turned on, the internal battery
charge level briefly appears on the “
battery shaped
”
gauge in the bottom left corner of the display.
VERIFYING CONFIGURATION
P
HARM
G
UARD
®
M
EDICATION
S
AFETY
HOSPITAL
VERSION
V6.0.0
M001234
CONFIG VERSION:
01.05.00002
CONFIG CRC: 4A4F
5
SECONDS REMAINING
©
99-2013
100%
When running an infusion on internal battery,
this “
battery shaped
” gauge also appears briefly in
the upper right corner of the display at the start of
delivery.
Pvd
0.18
m
L
rate
5
m
L/
Hr
LOCK
N
CHG RATE
CLEAR TVD
100%
B-D
60
m
L
Briefly display this “
battery shaped
” gauge at any time
while the pump is infusing on battery by pressing
â
.
Battery guidelines
Power & Battery Limits
When to charge
the battery
When the “
Battery Low
”
message displays, plug the
pump into a power outlet for
recharging. The battery indicator
remains on continuously when
charging. If the pump IS plugged
into an outlet and fails to charge
the battery, or the pump has been
stored for longer than two months
without charging, contact Service
& Repair for information on how
to test and/or service the battery.
When you
can
run on battery
The internal battery is intended
to allow infusions of medications
and fluids when an external
power outlet is not conveniently
available or when the patient is
being moved.
When you
cannot
run on
battery
When the battery becomes fully
depleted, all infusion stops, and
a “
Battery Depleted
” alarm
begins sounding. For a fully
depleted battery, plug in the AC
power cord, turn the pump off to
clear the alarm, and then turn the
pump back on. Leave the pump
plugged in for at least 16 hours to
completely recharge the battery.
•
See “General System Alarms and Alerts” table to
identify battery message status
.
Summary of Contents for Medfusion 3500