Medfusion® Model 3500 pump V6
Operator’s Manual
for each possible delivery mode with all functions
enabled.
Unloading the syringe
WARNING:
Turn off fluid flow
. Always turn
off fluid flow to patient in tubing via clamp or
stopcock before loading or unloading a syringe.
Uncontrolled fluid flow can occur when infusion
set is not clamped or turned off resulting in
inappropriate delivery rate, and may cause serious
injury and/or death.
Unload a syringe exactly the opposite way as it was
loaded. Just use the following steps:
1. Press
á
to stop delivery.
2. Clamp or close infusion line from the
syringe to the patient.
3. Unthread the tubing from the Tubing
Holders on the top left side of the pump.
4. Lift and swivel the syringe Barrel Clamp out
of the way.
5. Squeeze the Plunger Release Lever on the
syringe plunger and remove the syringe.