EMC information for the Medfusion® Model 3500 pump V6
Electromagnetic Compliance (EMC) precautions
are necessary when dealing with the Medfusion
®
Model 3500 pump V6. Accessible pins of connectors
identified with the ESD warning symbol near them
should not be touched with the fingers or hand-
held tool, unless proper precautionary measures are
followed. These measures include:
1) Discharging one’s body to earth or to a large metal
object.
2) Bonding oneself by means of an ESD wrist strap
to earth.
It is recommended that all staff, including clinical
and biomedical staff, receive training in the meaning
of the ESD warning symbol, and on the above
precautionary measures.
This training should also include an introduction to
the physics of electrostatic discharge, the voltage levels
that can occur in normal practice, and the damage
that can be done to electronic components if they
are touched by an operator who is electrostatically
charged. Further, an explanation should be given
of the methods to prevent build-up of electrostatic
charge, and how and why to discharge one’s body to
earth, or to bond oneself by means of a wrist strap to
earth prior to making a connection.
Use of cables
WARNINGS:
• Use of cables other than as listed below may
result in increased EMC emissions or decreased
immunity of the pump.
• Use of the listed cables with equipment other
than those specified may result in increased
EMC emissions or decreased immunity of the
equipment or system.
• The pump should not be used adjacent to or
stacked with other equipment. If adjacent
or stacked use is necessary, the user should
verify normal operation of the pump in the
configuration in which it is to be used.
This pump has been tested for EMC with the
following cable:
• Smiths Medical G6000329 AC power cord
(USA)
Stacked equipment
WARNING:
The pump should not be used adjacent
to or stacked with other equipment. If adjacent or
stacked use is necessary, the user should verify normal
operation of the pump in the configuration in which
it is to be used.
Medical Electrical Equipment needs special
precautions regarding EMC and needs to be
installed and put into service according to the EMC
information provided in this manual.
Portable and mobile RF communication equipment
can affect Medical Electrical Equipment.