89
Medfusion® Model 3500 pump V6
Operator’s Manual
Flush feature
Some medications and infusions may require
flushing the line to remove any residual medication
or infusate. Using PharmGuard
®
Toolbox 2, this
feature can be enabled for certain profiles, facilitating
the pump to automatically prompt flushing the
tubing when an infusion is complete. The pump may
prompt to set a delivery time for the flush, or only to
specify a volume (depending on how the feature is
enabled). This feature is only available if enabled for
intermittent (timed) infusions (for example, Dose/
kg/Time, Volume/Time, mL/Kg/Time, etc.), and is
not available for continuous infusions. Depending
on the pump’s custom configuration, a volume
limit may have been set in the drug program using
PharmGuard
®
Toolbox 2.
Volume only
1. When the programmed infusion is
complete or the
Syringe Empty
alarm
activates and the pump alarms, press
í
.
2. The
Begin Flush Setup
screen appears.
If flush is not desired at this point, press
No
(the pump will return to the
Begin
Infusion
screen). To begin the flush setup,
press
Yes
.
3. Remove the syringe at the end of the
infusion when prompted and replace with
the flush syringe. Use the
number
key to
select the syringe model.
4. Load the syringe onto the pump, verify the
correct size is shown, then press
Enter
.
5. Use the
number
keys to enter the desired
amount of flushing infusate, then press
Enter
. Volume is limited by the size of the
syringe being used, and may also be limited
by any custom configuration settings.
6. To begin flush delivery, press
Confirm
.
OR
• Press
Chg Vol
to return to the
Flush
Volume
screen and change the volume.
OR
• Press
Main Menu
(or
Quick Library
) to
skip the flush delivery and return to the
Main
menu.
OPTIONS
CLEAR TVD
INFUSION COMPLETE
VOLUME
4
m
L
TVD
4
m
L
TIME
00:10:00
TIME REMAINING
00:00:00
RATE
24
m
L/
Hr
MAIN MENU
NO
BEGIN FLUSH SETUP?
YES
SELECT SYRINGE MODEL
1
B-D
2
MONOJECT
3
TERUMO
ENTER
LOAD SYRINGE - PRESS ENTER WHEN READY
B-D
5
m
L
BACKSPACE
ENTER
FLUSH VOLUME TO INFUSE - PRESS ENTER TO CONTINUE
HIGH: 3
LOW: 0.0333
CLEAR
m
L
CHG VOL
CONFIRM
CONFIRM FLUSH SETTINGS?
MAIN MENU
VOLUME
1.5
m
L
Pvd
4
m
L
TIME
00:03:45
RATE
24
m
L/
Hr