Medfusion® Model 3500 pump V6
Operator’s Manual
Occlusion Alarm
For 1 mL Syringes, the occlusion limit is fixed at
Very High
.
Very High
(1 mL syringe only, not selectable): 50 psi (2585 mmHg) (345 kPa)
For Syringes with a volume greater than 1 mL, the user can set the occlusion limit
from Very Low to High.
High:
16 psi (827 mmHg) (110 kPa)
Normal:
12 psi (621 mmHg) (83 kPa)
Low:
8 psi (413 mmHg) (55 kPa)
Very Low:
4 psi (206 mmHg) (28 kPa)
Maximum volume
infused under single
fault conditions
1/100
th
of the syringe size by volume
Maximum delivery rate
during bolus
The configured maximum rate at which a bolus or loading dose can be delivered is:
Syringe size ×5mL/hr, Syringe size ×10mL/hr or Syringe size Max mL/hr (see page
133 for further explanation)
Maximum delivery rate
during Prime
The priming delivery rate is 300 mL/hr or the default maximum bolus rate for the
syringe size being used (Syringe size ×5mL/hr, Syringe size ×10mL/hr or Syringe
size Max mL/hr), whichever is lower (see page 133, Maximum Bolus Rate, for
further explanation)
Infusion rate range:
0.01 mL/hr to 1130 mL/hr in increments of 0.01/0.1 mL/hr (depending on syringe
size/rate). Refer to syringe tables on page 129 for Minimum Recommended Rate.
Bolus volume range:
Volume is limited to the maximum fill volume of the current syringe size, and
minimum volume is 1/600
th
the syringe size
Note:
The Minimum Recommended Bolus Volume is 10% of syringe or 1 mL,
whichever is smaller. Bolus volumes delivered that are less than the recommended
minimum bolus volume for the syringe used may result in delivery inaccuracies.
Delivery modes (user
customized)
Continuous Delivery Modes:
Dose / min, hr, day
Dose / kg / min, hour, day
Dose / m2 / min, hr, day
mL / hr
mL / kg / min, hr
Intermittent Delivery Modes:
Dose / kg / time
Dose / m2 / time
mL / kg / time
Volume / time
Dose / time
Intermittent volume / time
Delivery Options (user
customized)
Volume Limit
Bolus Dose
Loading Dose
KVO
Standby
Override Occlusion Limit
Periodic Callback
Toggle FlowSentry™
Delayed Start
Override Alarm Loudness
Toggle PVD/PDD
Disable/Enable Vol Empty Tone
Disable/Enable Near Empty Tone