63
Medfusion® Model 3500 pump V6
Operator’s Manual
CHG BOLUS
NO
BEGIN BOLUS DELIVERY?
BOLUS
2
MG/KG
VOLUME
8
m
L
TIME
01:00
MM:SS
RATE
480
m
L/
Hr
YES
totaL
boLUS
doSe
80
mg
PRESS CANCEL TO STOP BOLUS DELIVERY
DELIVERING
BOLUS
2
mg
/
kg
DELIVERED
3.0760
m
L
REMAINING
00:37
MM:SS
30.7600
MG
TOTAL BOLUS
80
MG
RATE
480
m
L/
Hr
CANCEL
CANCEL
– stops the bolus and returns to the normal
delivery rate and screen.
Bolus delivery
1. Press
è
(if necessary).
2. The
Begin Bolus Delivery?
screen
appears. To begin bolus delivery, press
Yes
.
If bolus delivery is not desired right now,
press
No
. To change the bolus amount,
press
CHG Bolus
and reprogram the
bolus (items 3 through 5 in the preceding
section).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
3. If
Yes
was pressed, the
Bolus Delivery
screen appears, with “
Delivering
”
and the drug program name alternately
appearing.
If
No
was pressed, the pump returns to
the delivery screen. Note that the bolus
information appears on the screen.
Pressing
è
will cause the
Begin Bolus
Delivery?
screen to reappear, where
bolus delivery can be started (see above).
LOCK
N
CHG DOSE
OPTIONS
CLEAR TVD
conc
10
mg
/
m
L
tvd
31.338
m
L
WeigHt
40
kg
kvo
boLUS
2
mg
/
mL
doSe
4
mg
/
kg
/
Hr
rate
16
m
L/
Hr
ED
k
etamine
10
mg
/
m
L
≥
CHG BOLUS
NO
BEGIN BOLUS DELIVERY?
YES
totaL
boLUS
doSe
80
mg
BOLUS
2
MG/KG
VOLUME
8
m
L
TIME
01:00
MM:SS
RATE
480
m
L/
Hr