Medfusion® Model 3500 pump V6
Operator’s Manual
Type
Definition and Remedy
Low priority
A
low-priority alarm
indicates any condition not requiring immediate operator
intervention.
Low-priority alarms
are announced with a continuous yellow indicator, a viewing screen
backlight that oscillates between bright and dim and a non-repeating audible signal.
During a low-priority alarm condition, all keys except
í
are locked.
You must silence
the alarm by pressing
í
before any other action can be taken
.
If the front panel controls are
locked
when a low-priority alarm occurs, the pump controls
do not
unlock.
Limit priority
A limit priority alarm provides user feedback related to interaction with the pump.
Note:
The audio part of the limit priority alarm is tied to “Key Click loudness”, which is
set in PharmGuard
®
Toolbox 2. If key click loudness is set to
Off
, there will be no audible
tone accompanying limit priority alarms (visual advisory messages will still appear in the
display). The volume will otherwise sound at the loudness level selected for key clicks -
Level 2 (quietest), Level 3, Level 4, or Level 5 (loudest).
A
limit-priority alarm
sounds a three tone, non-repeating audible signal and displays an
advisory message on screen for 3 seconds.
If the front panel controls are
locked
when a low-priority alarm occurs, the pump controls
do not
unlock.
Re-displaying an alarm message
Whenever an alarm is silenced, the alarm message
remains on the screen for approximately 3 seconds.
This message can be re-displayed for an additional 3
seconds by pressing
í
.