Medfusion® Model 3500 pump V6
Operator’s Manual
Volume limit
A
Volume Limit
is an
Option
available for
continuous
delivery modes
. On reaching the preset volume limit,
the pump stops – or changes to KVO if a keep vein
open rate has been programmed.
Volume limit is based on the amount of programmed
volume that is delivered.
Enable / Disable Volume Limit
: If
Volume Limit
is
not currently available as a dosing option, it can be
enabled through custom pump configuration (using
PharmGuard
®
Toolbox 2).
Setting volume limit
A volume limit can be set either before starting
an infusion or by pausing the pump during an
infusion. Volume limit is found in the
Options
menu, which can be accessed from the
Begin
Infusion
screen.
1. With the pump stopped, press
Options
.
2. Use the
number
key to select
Volume
Limit
.
3. Use the
number
keys to set the volume
limit, then press
Enter
.
Note:
Volume is limited to 500 mL regardless of
syringe size in use, as some infusions may require
multiple syringes.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
RATE
3.5
mL/hr
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
STANDBY
5
DELAYED START
2
VOLUME LIMIT
6
LOADING DOSE
3
OVERRIDE OCCL LIMIT
7
BOLUS DOSE
4
OVERRIDE ALARM LOUDNESS
8
KVO
BACKSPACE
ENTER
ENTER VOLUME LIMIT - PRESS ENTER TO CONTINUE
HIGH: 500
LOW: 0
CLEAR
m
L
50
DISABLE
RATE
3.5
m
L/
Hr
Disable
– turns Volume Limit off.
CLEAR
– deletes an
entry.
Backspace
– erases a character.