Medfusion® Model 3500 pump V6
Operator’s Manual
Delivering a loading dose
1. When ready, press
É
.
2. The
Begin Loading Dose Delivery?
screen appears. Depending on which mode
the pump is delivering in, the screen is
slightly different.
3. To begin delivering the loading dose, press
Yes
.
OR
• Press
No
to skip the loading dose and
return to the normal delivery screen.
OR
• Press
Chg Dose
to change the loading
dose settings.
• If enabled in PharmGuard
®
Toolbox 2
and if the loading dose settings have been
programmed in body weight, the key
labeled
Chg In mL
will allow
the dose to be
changed in mL directly (if enabled).
Note:
If no keys are pressed for 30 seconds, the
User Callback alarm will sound.
4. If
Yes
was pressed, the
Loading Dose
Delivery
screen appears and the infusion
begins, with “
Delivering
” and the drug
program name alternately appearing.
The pump delivers the loading dose then
automatically switches to the normal delivery rate
and screen. The loading dose volume delivered is
automatically added to the Total Volume Delivered
(TVD) as well as the Programmed Volume
Delivered (PVD), or Programmed Dose Delivered
(PDD) for that infusion.
PRESS CANCEL TO STOP LOADING DOSE DELIVERY
DELIVERING
LOAD
5
MG/KG DELIVERED
0.9550
m
L
REMAINING
02:42
MIN:SS
19.1000
MCG
TOTAL LOAD
200
MG
RATE
200
m
L/
Hr
CANCEL
CANCEL
– stops the loading dose and returns to the
normal delivery rate and screen.
CHG DOSE
NO
BEGIN LOADING DOSE DELIVERY?
LOAD
5
MG/KG
TIME
03:00
MM:SS
RATE
200
m
L/
Hr
YES
totaL
Loading
doSe
200
mg
CHG DOSE
NO
BEGIN LOADING DOSE DELIVERY?
YES
totaL
Loading
doSe
200
mg
LOAD
5
MG/KG
TIME
03:00
MM:SS
RATE
200
m
L/
Hr