Medfusion® Model 3500 pump V6
Operator’s Manual
Standby
Standby
is an
Option
that allows the pump to remain
at the
Begin Infusion
screen without the
User
Callback
alarm sounding for the Standby period.
• The pump
does not automatically start
an
infusion at the end of the standby.
• Instead, the pump triggers an alarm at the end of
the standby period.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
STANDBY
5
DELAYED START
2
VOLUME LIMIT
6
LOADING DOSE
3
OVERRIDE OCCL LIMIT
7
BOLUS DOSE
4
OVERRIDE ALARM LOUDNESS
8
KVO
BACKSPACE
ENTER
ENTER STANDBY TIME - PRESS ENTER TO CONTINUE
HIGH: 24:00:00
LOW: 00:00:00
CLEAR
HR : MIN : SEC
01:00
:00
OPTIONS
CLEAR TOTALS
STANDBY TIME REMAINING 00:59:50
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
Enable / Disable Standby
: The
Standby
option can
be enabled / disabled through custom configuration
(using PharmGuard
®
Toolbox 2).
Programming standby
1. When the pump is paused at the
Begin
Infusion
screen, press
Options
.
2. Use the
number
keys to choose
Standby
.
3. Set the Standby time in hours & minutes.
The maximum Standby setting is 24 hours.
The default Standby time is 1:00 Hr. When
the Standby time is correct, press
Enter
to accept the setting and go to the
Begin
Infusion
screen.
4. The Standby timer then appears in the upper
right corner on screen “
counting down
”
alternately with “
Press <
«
> To Begin
Infusion
” and the Drug Program name.
The
User Callback
alarm is disabled for
this interval. Standby continues until the
timer has counted to zero. Standby can be
discontinued at any time by pressing
É
,
when
Main Menu
is pressed to return to the
Main
menu, or by setting Standby to zero.
5. If this is a new infusion, prior to connecting
to the patient press
è
to initiate the
priming sequence (page 58).