87
Medfusion® Model 3500 pump V6
Operator’s Manual
Infusion Delivery
Starting & stopping infusion
delivery
An infusion delivery can be started at any time after
a delivery mode has been set up. An infusion can be
started from a pause, standby, or delayed start screen.
Prior to connecting the tubing to the patient and
starting an infusion, press
è
to initiate the priming
sequence
(page 58).
Start delivery from pause
Whenever beginning an infusion setup and
progressing all the way through the setup levels,
the pump
pauses
at the
Begin Infusion
screen.
1. Program an infusion.
2. Verify all values are correct.
3. Press
É
.
4. The pump begins the infusion.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CONC
250
mcg
/
m
L
PVD
0.00
m
L
WEIGHT
5
KG
DOSE
1
MCG/KG/MIN RATE
1.2
m
L/
Hr
CHG DOSE
MAIN MENU
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that prescribed
will cause inappropriate delivery, which can result in
serious injury or death.