79
Medfusion® Model 3500 pump V6
Operator’s Manual
Delayed start
Delayed Start
is an
Option
that allows the pump to
remain at the
Begin Infusion
screen without the
User Callback
alarm sounding for the specified period,
then
automatically
starts the infusion at the end of
the Delayed Start period.
Programming delayed start
1. When the pump is paused at the
Begin
Infusion
screen, press
Options
.
2. Use the
number
keys to choose
Delayed
Start
.
3. Set the Delayed Start time in hours &
minutes. The maximum Delayed Start
setting is 6 hours. The default Delayed Start
time is 1:00 Hr. When Delayed Start is
correct, press
Enter
to accept the setting and
go to the
Begin Infusion
screen.
4. The Delayed Start timer then appears in
the upper right corner on screen “
counting
down
” alternately with “
Press <
«
> Key
To Start Infusion Immediately
” and
the Drug Program name. The
User Callback
alarm is disabled for this interval. Delayed
Start continues until the timer has counted
to zero. Delayed Start can be discontinued
at any time by pressing
É
, or by setting
Delayed Start to zero.
5. If this is a new infusion, prior to connecting
to the patient press
è
to initiate the
priming sequence (page 58).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
6. When the counter reaches zero, the pump
automatically begins the infusion
.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
STANDBY
5
DELAYED START
2
VOLUME LIMIT
6
LOADING DOSE
3
OVERRIDE OCCL LIMIT
7
BOLUS DOSE
4
OVERRIDE ALARM LOUDNESS
8
KVO
BACKSPACE
ENTER
ENTER DELAYED START TIME - PRESS ENTER TO CONTINUE
HIGH: 06:00:00
LOW: 00:00:00
CLEAR
HR : MIN : SEC
01:00
:00
OPTIONS
CLEAR TOTALS
TIME UNTIL START OF INFUSION 00:59:50
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
Enable / Disable Delayed Start
: The
Delayed Start
option can be enabled / disabled through custom
configuration (using PharmGuard
®
Toolbox 2).