Medfusion® Model 3500 pump V6
Operator’s Manual
Keypad lock
The keypad can be locked during delivery. The lock
feature is provided to make the delivery settings
tamper resistant.
Once the keypad is locked, a setting cannot
be modified until the keypad is unlocked. The
Medfusion
®
Model 3500 pump V6 offers a
programmable unlock code option set through
custom configuration.
Any high-priority alarm
automatically unlocks the front panel.
1. Press
Lock
to lock the front panel controls.
The
Lock indicator on the front panel
lights.
2. If any front panel keys other than
Unlock
are pressed, a “
Pump is Locked
” message
appears on screen.
3. Press the
Unlock
softkey to unlock the
front panel. If an Unlock Passcode has been
set, a prompt to supply the passcode before
the front panel will unlock is displayed.
BACKSPACE
ENTER
ENTER UNLOCK PASSCODE - PRESS ENTER TO CONTINUE
HIGH: 9999
LOW: 0
CLEAR
PVD
0.025
m
L
RATE
4.5
m
L/
Hr
LOCK
N
CHG RATE
OPTIONS
CLEAR TVD
100%
B-D
60
m
L
OPTIONS
CLEAR TVD
PUMP IS LOCKED
PVD
0.030
m
L
RATE
4.5
m
L/
Hr
CHG RATE
UNLOCK
N