47
Medfusion® Model 3500 pump V6
Operator’s Manual
Volume / weight infusions
The following volume/weight modes may be available
on a pump: mL/kg/min and mL/kg/hr.
Volume/weight infusion rate calculations allow
programming by:
•
Patient’s weight
•
Volume of drug per kilogram of weight
•
Volume limit, KVO, loading dose, or bolus
dose may also be programmed (these options
are only available if they have been enabled
in the custom program using PharmGuard
®
Toolbox 2)
The pump calculates the correct flow rate in mL/
hr to achieve the required drug dose. To program
a volume/weight infusion, at the selection menu,
choose either
mL/Kg/Min
or
mL/Kg/Hr
(or choose
a drug program that uses the volume/weight mode
infusion type). Choose the syringe model and size,
and load the syringe as instructed on page 33.
1. Use the
number
keys to set the patient’s
weight in kilograms. Weight is limited to
the indicated range. Press
Enter
to accept
the patient weight setting.
The pump may be customized with an
additional patient weight confirmation
screen. Re-enter the patient weight, then
press
Enter
to accept the setting.
2. At the
Volume/Kilogram
screen, press
the
number
keys to set the volume per
weight setting. Range limits are determined
by the installed syringe size, patient
weight, and/or the custom setup in the
Configuration. When the setting is correct
press
Enter
to accept it.
3. If desired (and if they are enabled), program
a Bolus Dose, Loading Dose or Volume
Limit (see Options, page 60).
4. At the
Begin Infusion
screen all values
are set and the pump pauses (the pump may
be configured with additional confirmation
screens, requiring confirmation of all
settings again before beginning the
infusion).
5. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
BACKSPACE
ENTER
ENTER PATIENT WEIGHT - PRESS ENTER TO CONTINUE
HIGH: 250
LOW: 0.25
CLEAR
KG
5
BACKSPACE
ENTER
RE-ENTER PATIENT WEIGHT - PRESS ENTER TO CONTINUE
HIGH: 250
LOW: 0.25
CLEAR
KG
BACKSPACE
ENTER
ENTER VOLUME / KILOGRAM
-
PRESS ENTER TO CONTINUE
HIGH: 0.3267
LOW: 0.0002
CLEAR
m
L /
kg
/
Hr
0.
3
Summary of Contents for Medfusion 3500