90
Medfusion® Model 3500 pump V6
Operator’s Manual
Note:
If no keys are pressed for 30 seconds, the
User Callback
alarm will sound.
7. When
Confirm
is pressed, the flush
infusion begins delivering, with
“
Flushing
” and the drug program name
alternately appearing. The flush is calculated
to deliver at the same rate as the previous
infusion.
To cancel or stop the delivery, press
á
.
If flush delivery is cancelled/stopped, or
should an alarm interrupt flush delivery, It
can be restarted (see below).
8. When the flush infusion has been delivered,
delivery stops and the pump alarms. Press
í
to silence the alarm, then press
Main
Menu
.
PRESS <
> TO STOP FLUSH DELIVERY
FLUSH VOLUME
1.5
m
L
FLUSHED
0.140
m
L
TIME
00:03:45
REMAINING
00:03:22
RATE
24
m
L/
Hr
FLUSHING
CONTINUE SAME FLUSH?
FLUSH VOLUME
1.5
m
L
FLUSHED
0.151
m
L
TIME
00:03:45
REMAINING
00:03:21
RATE
24
m
L/
Hr
YES
Pvd
4.151
m
L
NO
FLUSH VOLUME
1.5
m
L
FLUSHED
0.151
m
L
TIME
00:03:45
REMAINING
00:03:21
RATE
24
m
L/
Hr
CANCEL
OCCLUSION - CHECK INFUSION LINE
Continuing an interrupted flush infusion
Should a
High Priority
alarm occur during flush
infusion delivery, or if
á
is pressed to interrupt a
flush infusion, delivery will stop. Either continue
the flush from where it left off, restart a new flush,
or cancel the flush.
1. If necessary, press
í
to silence the alarm.
Troubleshoot the alarm condition.
2. Press
É
.
3. To continue the flush infusion from
where it left off (delivering only what was
remaining), press
Yes
.
Press
No
to cancel the remainder of the
flush infusion and return to the
Main
menu.
FLUSH STOPPED - PRESS < « > KEY TO CONTINUE
FLUSH VOLUME
1.5
m
L
TIME
00:03:45
REMAINING
00:03:22
RATE
24
m
L/
Hr
YES
Pvd
4.140
m
L
NO
INFUSION COMPLETE
VOLUME
1.5
m
L
TIME
00:03:45
REMAINING
00:00:00
RATE
24
m
L/
Hr
MAIN MENU
Pvd
5.5
m
L
FLUSH STOPPED - PRESS < « > KEY TO CONTINUE
FLUSH VOLUME
1.5
m
L
TIME
00:03:45
REMAINING
00:03:21
RATE
24
m
L/
Hr
YES
Pvd
4.151
m
L
NO
Summary of Contents for Medfusion 3500