118
Medfusion® Model 3500 pump V6
Operator’s Manual
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Medfusion® Syringe Pump Model 3500 V6 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Medfusion® Syringe Pump Model 3500 V6 should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF Emissions
CISPR 11
Group 1
The Medfusion® Syringe Pump Model 3500 V6 uses
RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment RF emissions.
RF emissions
CISPR 11
Class A
The Medfusion® Syringe Pump Model 3500 V6 is
suitable for use in all establishments other than
domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Not Applicable
Compliance using:
1. AC power cord length of 2.4 m (8 feet).
2. Rechargeable NiMH battery pack.
WARNING:
The pump should not be used adjacent
to or stacked with other equipment. If adjacent or
stacked use is necessary, the user should verify normal
operation of the pump in the configuration in which
it is to be used.