Sirona Dental Systems GmbH
11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.3.8
S017: Test step 12
Enabling/disabling the preview image
* Factory setting
1. Call service routine S017.12 [ → 125].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
IMPORTANT
The preview image is automatically disabled for volume exposures,
regardless of the setting in service routine S017.12.
Selection field
Code
Function
01
00
Preview image
disabled
01
Preview image
enabled*