13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
13.2.2.4
Shading (5x5)
✔
The geometry phantom
is not
in the pan bite block holder on the unit.
1. In the structure tree, under
"3D Adjustment/calibration"
, click on the
Shading element (S010.18).
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (85kV/18 mAs) [ → 206].
A message is displayed in the action window to indicate whether
adjustment has been successful and the traffic light symbol in
front of the Shading element in the structure tree responds
accordingly:
Calibration not OK = traffic light stays red
Calibration OK = traffic light turns yellow
4. Repeat the procedure starting at Step 2 until calibration is successful.
5. When calibration is OK, save the values.
The traffic light symbol in front of the Shading element in the
structure tree changes color to green (with check mark). This
indicates that the calibration settings are valid.
6. Continue with the next stage of the calibration procedure [ → 254].