12.8 Checking the laser light localizers
Installation Manual ORTHOPHOS XG 3D / Ceph
Checking the MS laser: XG 3D
Checking the vertical laser beam (MS)
1. Affix a piece of white cardboard between the temple supports.
2. Rotate the mirror by pressing into the left depression (D) of the
toolbar.
3. Press the light localizer key (E) on the control panel.
Then light localizers are switched on.
The laser beam is displayed on the cardboard by a red line.
The vertical laser beam must be displayed in the center of the
forehead support and the bite block holder. If this is not the case,
adjust the laser light localizer.
4. Switch the light localizers off.