13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2
Adjustment and calibration via the "Unit
adjustment/calibration" menu
13.2.1
2D adjustment
13.2.1.1
Pan sensor adjustment
13.2.1.1.1
Automatic adjustment: Panoramic sensor (standard)
✔
The pan sensor is plugged into the pan slot on the unit.
✔
The OP XG needle phantom is in the pan bite block holder on the
1. Call the
"Adjustment/calibration"
2. In the structure tree, under
"2D Adjustment"
, click on the
"Pan -
Sensor Adjustment"
element.
The
"Pan - Sensor Adjustment"
menu is displayed in the action
area. The menu supports precision adjustment and coarse
adjustment (precision adjustment is pre-selected). Perform a
precision adjustment first. In most cases, previous coarse
adjustment is not necessary.
3. Make SIDEXIS XG ready for exposure. [ → 206]
4. Take an exposure (60 kV / 3 mA; 0.6 s) [ → 206].
SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
"Pan - Sensor Adjustment"
menu.
The exposure is displayed in the exposure window in SIDEXIS
XG.
IMPORTANT
After every adjustment or calibration of the unit, the reference values for
the constancy measurement must be recalculated and entered in the
"Test results" form, "Reference value" column.