63 03 452 D3352
156
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4.4
S018: Test step 5
Setting the minimum travel height
Setting the minimum travel height is possible only for a unit height that is
below the lower correction switch level (< position value of 1500)!
Minimum travel height: XG 3D
1. In user mode, move the unit to the required minimum travel height by
pressing the Up/Down keys (A).
2. Call service routine S018.5 [ → 125].
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
3. To save the minimum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
4. Exit the service routine [ → 129].
The limitation of the minimum travel height is purely software based. The
lower limit switch is not mechanically adapted to the new minimum travel
height!