63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
129
Sirona Dental Systems GmbH
11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.2.4
Saving parameters
Once one or a number of parameters have been selected via a service
routine, the current selection must be saved so that it is applied in the unit.
Save Easypad parameter: GALILEOS / XG 3D / XG 3D ready
Example
You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔
The Memory key (C) lights up.
1. Touch the Memory key (C).
The R key (D) lights up.
2. Touch the R key (D).
The selected setting is saved to non-volatile memory.
11.10.2.5
Exiting the test step and service routine
Exit test step: GALILEOS / XG 3D / XG 3D ready
Touch the Service key (A) or the double arrow key (B) to go back to the
menu for selecting service routines.
Touch the double arrow key (B) in the service menu to go back to the
main menu.
Exception: Service routine S017
In service menu S017, touch the Service key (A) to go to the next test step
in the service routine.