11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.3.13
S017: Test step 18 (occlusal bite block only)
Activation of occlusal bite block function
When the occlusal bite block function is activated, all previously existing
calibration values are overwritten with default data (factory setting) where
applicable. Once the function has been activated, the unit sends error
message E1 11 18 to indicate the need for adjustment.
Following activation of the function, service routine S018.9 also has to be
run to check or execute the adjustment function.
A check also has to be made to ascertain whether the creep speed
required for operation has been configured on this unit. To do this, run
service routine S018.10.
Error message E1 11 18 disappears as soon as the two service routines
mentioned above have been performed.
Should the customer wish to set a nominal angle deviating from the
factory setting (7°), make the corresponding configuration settings via
service routine S018.7).
* Factory setting
1. Call service routine S017.18 [ → 149].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
Selection field
Code
Function
1
00
Occlusal bite block function
deactivated*
01
Occlusal bite block function activated
Summary of Contents for Orthophos XG 3D/Ceph