11 Initial startup
Sirona Dental Systems GmbH
11.7 Setting up the X-ray component via the SIXABCon program
Installation Manual ORTHOPHOS XG 3D / Ceph
Boot service
✔
The element
"Boot service"
is selected in the structure tree.
1. Switch the unit on.
The unit is detected in the network and the values for the IP
address, subnet mask and standard gateway are automatically
imported to the input fields of the menu.
2. If you want to, you can enter an individual IP address in the
"IP
address"
input field and then click into one of the other input fields
with the cursor.
Default values suitable for the IP address are automatically
entered in the
"Subnet mask"
and
"Standard Gateway"
input
fields.
3. You can either overwrite the default values in the input fields or leave
them unchanged, depending on the network involved.
IMPORTANT
Automatic addition and setup of the X-ray component via the boot
service functions properly only if the unit is started with the default
factory-set IP address.
If the IP address already has been changed, reset it to the factory setting
via service routine S037.2 or use the elements
"Automatic search"
or
"Add component manually"
(see section on "Alternatives to the boot
service".
IMPORTANT
If the unit is not automatically detected in the network, it is possible to
manually assign a static IP address via service routine S037.4.
IMPORTANT
Each unit must be assigned a unique IP address in the network.