13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1 General information about unit adjustment and calibration
båÖäáëÜ
13.1.1
Displays and help messages during adjustment/
calibration
The most frequent help and status messages during calibration are listed
below.
Help messages
H3 01: Move unit to starting position, press the R key.
H3 07 XG 3D / 3D ready
H3 07: Remove the occlusal bite block.
H3 21: Close the door.
H401/402 XG 3D
H4 02: Plug the sensor into the ceph slot (sensor missing).
H4 03:SIDEXIS XG is not ready for exposure, make unit ready for
exposure.
H406/404 XG 3D
H4 04:Plug the sensor into the ceph slot (incorrect sensor plugged in).
H4 06: Move cephalometer to starting position, press the R key.
H501/502 XG 3D
H5 01: Turn cephalometer into lateral position.
H5 02: Turn cephalometer into AP or PA position.
H610 XG 3D
H6 10: Wait until preview image is acknowledged.
Status messages
XG 3D status messages
●
"Ready for exposure"
●
"Exposure not possible"
●
"Please wait"
●
"Ready for exposure in XXs"
●
"X RAY Active!"
If error message E1 11 20 is displayed on the control panel and/or the
remote control during the calibration process, this does not necessarily
indicate an equipment error. This error message only indicates that the
adjustment or calibration data of the unit is incomplete at this point.
Acknowledge the error message with the R key, if applicable, and
continue the adjustment or calibration procedure.
For assistance with other help messages or error messages displayed
during the adjustment or calibration process, please refer to the section
of these instructions entitled Messages.