63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
15
Sirona Dental Systems GmbH
3 Unit description
Installation Manual ORTHOPHOS XG 3D / Ceph
3.2 Dimensions/Space requirements
båÖäáëÜ
3.2
Dimensions/Space requirements
3.2.1
Front view
Front view: XG 3D
These dimensions apply to installation of the unit without the floor stand.
Installation with the floor stand results in an additional 30 mm (1 1/4")
increase of all height dimensions (see also separate Installation
Requirements document).
NOTICE
Unit can travel to the ceiling.
The minimum ceiling height should be 2.10 m (82 3/4"). If the ceiling
height is lower than 2.27 m (89 1/2") (max. travel height of 2.25 m (88 1/
2")), the travel height of the unit must be adjusted or limited prior to
startup via service routine S018.2.
IMPORTANT
The measurements on the drawings are specified in both mm and
inches.
ORTHOPHOS
402,5
15 7/8”
2249
88 1/2”
1950
76 3/4”
A
A
Alternative fastening if it is not possible to screw the unit onto the
floor. Order bracket separately.
Summary of Contents for Orthophos XG 3D/Ceph
Page 1: ...k W f j NMKOMNR loqelmelp ud Pa L b b Cover page...
Page 2: ......