13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
A message is displayed in the action window to indicate that the
values have been successfully saved.
The traffic light symbol in front of the
"Sensor "
element in the
structure tree changes color to green (with check mark). This
indicates that the adjustment settings are valid.
9. Remove the geometry phantom from the pan bite block holder on the
unit.
10. Continue with the next stage of the adjustment procedure [ → 250].