12 Startup, measurements and tests (for USA/Canada only)
Sirona Dental Systems GmbH
12.8 Checking the laser light localizers
Installation Manual ORTHOPHOS XG 3D / Ceph
5. Move the lower horizontal laser beam by moving the bottom slider (D)
into the lower area of the 3D sensor area.
The laser beam must be parallel to the bottom border marking of
the 3D sensor area. If this is not the case, adjust the double laser
light localizer.
Minimum and maximum distances apply between the two laser
localizers. If the action of moving the slide controls causes these
distances to be undershot or exceeded, the slide controls must
act in tandem with one another. In other words, as soon as the
minimum or maximum distance is undershot or exceeded,
whenever the top control moves, the bottom control must move
along with it.
If the dual laser light localizer does not respond as required,
replace it.
6. Switch the light localizers off.