11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.3.7
S017: Test step 9
Activate/deactivate operation with board DX41
This service routine is used to configure operation of the system with or
without module DX41. This configuration is necessary for software
updates and module replacement of board DX11 if systems with or
without board DX41 should be supported.
* Factory setting
XG 3D / XG 3D ready
1. Call service routine S017.9 [ → 125].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required code in selection field 1 (see
table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
Selection field
Code
Function
1
00
Board DX41 inactive*
01
Board DX41 active