13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
4. Repeat the procedure starting at Step 2 until the value in the menu's
text box is equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce a value equal
to zero, calculate the adjustment value manually and overwrite the
value in the menu's text box with this value [ → 240].
5. If adjustment is OK (the adjustment value in the text box is equal to
zero), save the value [ → 206].
6. Continue with the next stage of the adjustment procedure [ → 248].
13.2.1.7.2
Manual adjustment: Ceph - Fixed point of rotation for QuickShot
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment value
calculated automatically by SIDEXIS XG in the
"Ceph - Fixed point of
rotation Quickshot"
menu is overwritten.
1. Start the
"Ceph - Fixed point of rotation Quickshot"
adjustment
procedure as described in the chapter entitled "Automatic
adjustment: Ceph - QuickShot fixed point of rotation
A
Adjustment OK
● The exposed image section must lie
centered and straight in the image
section as well as inside the
superimposed auxiliary lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
B
Adjustment not OK
Summary of Contents for Orthophos XG 3D/Ceph