63 03 452 D3352
130
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.3
S017: Configuration service
S017 XG 3D
* SR=service routine, ** SA=security access
11.10.3.1
S017: Test step 2
S017_2 XG 3D / 3D ready
Configuring the hardware version
* Factory setting
SR*
SA**
Function
S017
Unit configuration
S017.2
Yes
Configuring the hardware version
S017.3
Yes
Enter the country group code
S017.4
Yes
Select a language
S017.5
Yes
Select a language set
S017.6
Yes
Enable/disable the remote control
S017.8
Yes
Selecting the kV/mA level series
S017.12
Yes
Enabling/disabling the preview image
S017.13
Yes
Enable/disable the welcome screen
S017.14
Yes
Enable/disable certain lines of the welcome screen
S017.15
Yes
Activate/deactivate the acoustic signal for end of exposure
S017.17
Yes
Trigger save/restore function for DX61 data
S017.18
Yes
Activation of the occlusal bite block function (for occlusal bite block only)
NOTICE! This service routine may be selected only for units with an
occlusal bite block; otherwise, the unit will malfunction!
S017.21
Yes
Configuration of angle pre-selection (temporomandibular joint) for the
TM1 and TM2 user programs
Selection field
Code
Function
1
0001
Panoramic digital
0003
Panoramic digital, ceph left digital
0005
Panoramic digital, ceph right digital
0201
XG 3D + panoramic digital*
0203
XG 3D +
panoramic digital, ceph left digital
0205
XG 3D +
panoramic digital, ceph right digital