13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1 General information about unit adjustment and calibration
båÖäáëÜ
13.1.2.2.2
Coarse and precision adjustment
The
"Pan - Sensor Adjustment"
,
"Pan - Diaphragm"
, and
"Ceph - Fixed
point of rotation"
menus support coarse and precision adjustment
(precision adjustment is pre-selected). Always try to use precision
adjustment first when adjusting the unit. In most cases, previous coarse
adjustment is not necessary. If a coarse adjustment is required, you are
informed via a message window.
To perform coarse adjustment, uncheck the
"Precision adjustment"
box
and complete the adjustment procedure in the same way as for precision
adjustment.
The correction procedure required for coarse adjustment is identical to
that for precision adjustment. The only difference between the two modes
is the size of the image field considered. Furthermore, there are fewer
auxiliary lines in the coarse adjustment mode.